2016年5月19日星期四

Extractive HCL laser gas analyzer

Extractive HCL laser gas analyzer

Principle
laser gas analyzer using a Tunable Diode Laser Absorption Spectroscopy (TDLAS), the realization of the non contact measurement in high temperature, high pressure, high dust and other harsh test conditions. High selective laser based on frequency, analyzer has good performance of anti interference. compared to the traditional analysis method, reducing the complex sampling system, flexible installation and easy to use, greatly reduce the manual installation and maintenance workload
Specifications
Gas
HCL
Working principle
Tunable Dioxide Laser Absorption Spectroscopy (TDLAS)
Measurement Range
0 to 50 ppm
Accuracy
+/- 2%FS
Repeatability
+/- 1%FS
Resolution
0.01ppm
Response Time t90
< 15 s
Refresh frequency
2s
Ambient Temperature
-20 to 55 °C
HCL gas sensor module heating temperature
190 °C
Sample gas max humidity
20%. abs
Sample gas flow
1.5L-3.0L/min
Input sample gas temperature
>=190
Output Interface
RS232
Analog output
420mA
Power supply
200V-240V AC / 50Hz-60Hz  (100-120V AC/ 50-60Hz optional)
Temperature control
190℃
Inlet and outlet connector
Ø6 mm
Dimensions
485*450*175mm(L*W*H)
Weight
About 10kgs