Portable Infrared Syngas Analyzer

Applications
coal chemical process
steel making process as blast furnace,    converter, coking, direct iron ore smelting reduction
syngas production from Biomass and coal gas gasification processes
Heat treatment
Measurement technology
•Infrared dual beam NDIR for CH4,CO2,CO,CnHm
•Thermal conductivity  TCD for H2
•Galvanic fuel cell ECD for O2
•Gas calorific value :calculated

Min/Max available range
•CH4,CO2,CO,H2:0-10%/0-100%
•CnHm:0-10%
•O2”0-5%/0-100%
•Intermediary ranges available without price increase
•Resolution: 0,01%
•Accuracy: ≤ 2% FS
•Repeatability: ≤ 1% FS
•Response time: NDIR/ECD < 10 sec / TCD < 15 sec
•Warm-up time: 800 seconds
•Gas conditions at inlet
•Quality: no dust, tar, oil, water vapor and/or moisture
•Pressure : 2 to 50 Kpa
Standard features
• Robust casing with stainless steel /composite material
• Temperature regulated enclosure for NDIR and TCD detector
• Large LCD display (320×240)
• Push button keyboard interface for configuration and calibration
• Integrated flow meter with needle valve
• External removable inline filter
• Internal battery (autonomy > 8h)
• Data  acquisition  function  (logging  rate  adjustable  from  3  to  99 seconds, capacity 1500 sets of up to  8 data, identification of up to 10 different sites and 100 measuring points)
• Internal gas sampling pump (nominal flow 1l/min)
• RS232 COM port
• Memory & real time data download software included
• Dimensions: W415 × D150 × H285 mm
•  Weight:  7 kg

Portable Infrared Syngas Analyzer



MeasurementCO, CO2, CH4, CnHm, O2, H2  + BTU index (gas calorific value)
CalculationHigh Heating Value or Low Heating Value in MJ/m3 or   kcal/m3
N2(Optional)
TechnologyCO, CO2, CH4, CnHm: proprietary dual beam NDIR detectors
O2  :   industrial electrochemical cell
H2   :   proprietary thermal conductivity detector
No effect of CO2 and CH4 on H2 detectorH2 reading is compensated for the interference effects of   the other gases measured
Resolution0,01%
Accuracy≤ ± 2% FS
Repeatability≤  2%
ZeroAuto-zeroing function via keyboard interface
Gas conditions at analyzer inletFlow:0.7~1.2L/min, internal  gas sampling  pump, external flow meter with needle valve
Pressure:20~100mbar
Quality: No dust, moisture, tar (external safety filter and optional gas washing   unit)
Operating conditionsTamb : 0-50°C / Pamb : 86 to 108 kPa / RH : 0-95% non condensing
Response time (T90)≤  15 sec
Warm-up time800 seconds (auto-zeroing cycle during the last 120   seconds)
Communication interfaceRS232 ( real time and memory data download  software included)
Power supplyExternal 220 VAC-50Hz
Internal with battery and charger; Autonomy of  > 4h with pump in operation
Data loggingUp to 1500 sets of data; Logging rate adjustable from 3 to   99 sec
Possibility to identify 10 different sites and up to 100   measuring points
DisplayLCD 320 x 240 display with back-lit function
Simultaneous indication of the 7 measures and units
CasingNew design moulded casing and shoulder trap (delivered with protective carrying bag)
CasingRobust casing in aluminium with cover and shoulder trap
Dimensions and weightGas analyser: W415 × D150 × H285 mm
Portable gas washing unit : L400 x  D120 x H 380 mm /  ± 3 kg (5 kg max with packaging)

2016年10月18日星期二

Extractive Oxygen Laser Gas Analyzer

Specifications
Gas
Range
Resolution
Repeatability
Linearity
Response   time
Drift
O2
0-10%volcan be customized
0.1%
≤1%FS
≤1%FS
≤2S
≤±1%FS
Warm up
30min
Digital   output
RS232/485
Analog output
4-20mA
Power supply
DC18-36V/20W or
AC100-240V/47-63Hz
Relay output
Load   capacity: AC/DC 4V/1A; Concentration limitation alarm, abnormal transmittance   alarm, abnormal laser temperature alarm
Optical path
1-12m
Flue gas temperature
≤1100
Flue gas pressure
Atmosphere pressure±5kPa
Dust content
10000mg/Nm3
Protection level
IP65
Explosion level
ExdIICT5optional
Flushing gas
Industrial   N2 gas >0.3MPa
Ambient temperature
-10~55
Ambient pressure
70kPa-120kPa

In-situ Oxygen Laser Gas Analyzer iLAS-100

Specifications
Gas
Range
Resolution
Repeatability
Linearity
Response   time
Drift
O2
0-10%volcan be customized
0.1%
1%FS
≤1%FS
≤2S
≤±1%FS
Warm up
30min
Digital   output
RS232/485
Analog output
4-20mA
Power supply
DC18-36V/20W or
AC100-240V/47-63Hz
Relay output
Load   capacity: AC/DC 4V/1A; Concentration limitation alarm, abnormal transmittance   alarm, abnormal laser temperature alarm
Optical path
1-12m
Flue gas temperature
≤1100
Flue gas pressure
Atmosphere pressure±5kPa
Dust content
10000mg/Nm3
Protection level
IP65
Explosion level
ExdIICT5optional
Flushing gas
Industrial   N2 gas >0.3MPa
Ambient temperature
-10~55
Ambient pressure
70kPa-120kPa

2016年10月9日星期日

Extractive NH3 Laser Gas Analyzer eLAS-100

Specifications
Specifications
Gas
Range
Resolution
Repeatability
Linearity
Response time
Drift
NH3
0-10ppm50ppm
0.1%
≤1%FS
≤1%FS
≤1S
≤±1%FS
Warm up
15min
Digital output
RS232/485
Analog   output
4-20mA   with max load 900Ω
Power   supply
AC100-240V/47-63Hz
120w
Relay output
Load capacity: AC/DC 4V/1A; Concentration   limitation alarm, abnormal transmittance alarm, abnormal laser temperature   alarm
Optical   path
1-3
Flue   gas temperature
≥180℃ no condense
Flue   gas pressure
Atmosphere   pressure±10kPa
Sample   gas flow
≤6L/min
Ambient   temperature
-10~55
Ambient   pressure
70kPa-120kPa
Dimension
I435X450X136mm  rack type 3U case
Weight
12kg


Online HCL Laser Gas Analyzer eLAS-200

Specifications
Performance   parameters
Gas
Range
Resolution
Repeatability
Linearity
Response   time
Drift
HCL
0-10ppm3000ppm
0.2ppm
1%FS
≤1%FS
≤1S
≤ ±1%FS
Function   parameters
Warm up time
15min
Digital   output
RS232/485
Analog output
4-20mA maximum load 900Ω
Power supply
AC100-240V/47-63Hz
120W
Relay output
Load   capacity: AC/DC 4V/1A; concentration overrun alarm, transmittance   and laser temperature abnormal alarm
Environmental   parameters
Optical path
3-10meters
Gas temperature
≥180non-condensate
Gas Pressure
Atmospheric pressure±5kPa
Sampling flow rate
6L/min
Environmental temp
-10~55
Environmental pressure
70kPa-120kPa
Size
3U enclosure 350X450X120mm
Weight
4.5kg