2016年12月21日星期三

Extractive Oxygen Laser Gas Analyzer

Specifications
Gas
Range
Resolution
Repeatability
Linearity
Response   time
Drift
O2
0-10%volcan be customized
0.1%
≤1%FS
≤1%FS
≤2S
≤±1%FS
Warm up
30min
Digital   output
RS232/485
Analog output
4-20mA
Power supply
DC18-36V/20W or
AC100-240V/47-63Hz
Relay output
Load   capacity: AC/DC 4V/1A; Concentration limitation alarm, abnormal transmittance   alarm, abnormal laser temperature alarm
Optical path
1-12m
Flue gas temperature
≤1100
Flue gas pressure
Atmosphere pressure±5kPa
Dust content
10000mg/Nm3
Protection level
IP65
Explosion level
ExdIICT5optional
Flushing gas
Industrial   N2 gas >0.3MPa
Ambient temperature
-10~55
Ambient pressure
70kPa-120kPa

Extractive Oxygen Laser Gas Analyzer


Features
  • high integration design, compact structure, simple installation and maintenance
  • no sample pretreatment needed, real-time measurement of oxygen concentration
  • single spectra scan, no interference from dust, tar and background gases
  • maintenance-free design
  • optional explosion-proof design
Applications

  •  
    process control/explosion pre-warning monitoring
  • environmental flue gas emission monitoring
  • boiler combustion control
  • air separation and gas production
  • aerospace, electric power, electronics, biological technology and scientific research institutions, etc
     

2016年12月6日星期二

Online Electrochemical H2S Gas Analyzer for CEMS Application


Operating Principle
3-electrode potentiostatic cell w/o potentiostat
Sensor Lifetime
2 years
Measurement Range
0 to 100 ppm
Maximum Overload
up to 250 ppm
Sensitivity
500 to 1,000 nA/ppm
Response Time t90
< 25 s
Drift
< 30 % signal/year
Zero Offset Current
< ± 1 ppm
Bias Voltage
+ 300 mV
Warm-up Time
8 h, when connected to Bias
Operating Temperature
10 to 45 °C(increased response time below 0 °C)
Pressure Range
750 to 1250 hPa
Humidity Range
30 to 60 % RH
Interferences
(without filter)
2.000 ppm CO: < 1 nA/ppm
1.000 ppm SO2: < 40 nA/ppm
800 ppm NO: < 400 nA/ppm
100 vol. % H2: none
Output Interface
RS232
Analog output
420mA
Working Temperature
1045℃
Relative Humidity
3060%
Pressure
750 to   1250 hPa
Power supply
200V-240V AC / 50Hz-60Hz  (100-120V AC/ 50-60Hz optional)
Temperature control
55℃
Calibration
Zero and span calibration
Dimensions
435*450*135 mm (L*W*H)
Weight
About 12kg

Online Electrochemical H2S Gas Analyzer for CEMS Application


Operating Principle
3-electrode potentiostatic cell w/o potentiostat
Sensor Lifetime
2 years
Measurement Range
0 to 100 ppm
Maximum Overload
up to 250 ppm
Sensitivity
500 to 1,000 nA/ppm
Response Time t90
< 25 s
Drift
< 30 % signal/year
Zero Offset Current
< ± 1 ppm
Bias Voltage
+ 300 mV
Warm-up Time
8 h, when connected to Bias
Operating Temperature
10 to 45 °C(increased response time below 0 °C)
Pressure Range
750 to 1250 hPa
Humidity Range
30 to 60 % RH
Interferences
(without filter)
2.000 ppm CO: < 1 nA/ppm
1.000 ppm SO2: < 40 nA/ppm
800 ppm NO: < 400 nA/ppm
100 vol. % H2: none
Output Interface
RS232
Analog output
420mA
Working Temperature
1045℃
Relative Humidity
3060%
Pressure
750 to   1250 hPa
Power supply
200V-240V AC / 50Hz-60Hz  (100-120V AC/ 50-60Hz optional)
Temperature control
55℃
Calibration
Zero and span calibration
Dimensions
435*450*135 mm (L*W*H)
Weight
About 12kg

2016年12月1日星期四

In-situ Oxygen Laser Gas Analyzer iLAS-100

Specifications
Gas
Range
Resolution
Repeatability
Linearity
Response   time
Drift
O2
0-10%volcan be customized
0.1%
1%FS
≤1%FS
≤2S
≤±1%FS
Warm up
30min
Digital   output
RS232/485
Analog output
4-20mA
Power supply
DC18-36V/20W or
AC100-240V/47-63Hz
Relay output
Load   capacity: AC/DC 4V/1A; Concentration limitation alarm, abnormal transmittance   alarm, abnormal laser temperature alarm
Optical path
1-12m
Flue gas temperature
≤1100
Flue gas pressure
Atmosphere pressure±5kPa
Dust content
10000mg/Nm3
Protection level
IP65
Explosion level
ExdIICT5optional
Flushing gas
Industrial   N2 gas >0.3MPa
Ambient temperature
-10~55
Ambient pressure
70kPa-120kPa

In-situ Oxygen Laser Gas Analyzer iLAS-100


Features
  • high integration design, compact structure, simple installation and maintenance
  • no sample pretreatment needed, real-time measurement of oxygen concentration
  • single spectra scan, no interference from dust, tar and background gases
  • maintenance-free design
  • optional explosion-proof design
Applications
  • process control/explosion pre-warning monitoring
  • environmental flue gas emission monitoring
  • boiler combustion control
  • air separation and gas production
  • aerospace, electric power, electronics, biological technology and scientific research institutions, etc

2016年11月24日星期四

Syngas Analyser Continuously Monitoring System

qBiomass and coal  gasification processes (up to 6 gases)  
O2 - CO2 - CO - CH4 - CnHm - H2  + BTU calculation
Measurement range can be customised according to customer reques
Gas name
Symbol
Measuring principle
Range1
Resolution
Accuracy
T90







Oxygen
O2
Fuel cell
0-25%
0.01%
 ±2% FS
< 15 s
Carbon dioxide
CO2
NDIR
0-30%
0.01%
 ±2% FS
< 10 s
Methane
CH4
NDIR
0-20%
0.01%
 ±2% FS
< 10 s
Carbon monoxide
CO
NDIR
0-40%
0.01%
 ±2% FS
< 10 s
Hydrogen
H2
Thermal conductivity
0-50%
0.01%
 ±2% FS
< 10 s
Hydrocarbons
CnHm
NDIR
0-10%
0.01%
 ±2% FS
< 10 s